Location:Home>Institute>Laboratory Building
Environmental Controllable Super-low Friction Tribo-systemHits [1579] Published :2013-11-08 11:42:28

INTRODUCTION:

Environmental Controllable Super-low Friction Tribo-system consists of stainless vacuum chamber, pin-on-disk tribotest system, gas environment control system (gas cleaner and gas hose), gas environment measurement system, and external signal collection. Through controlling gas environment and the measurement system, precise control on the gas species, humidity and oxygen content (lowest of 1 ppm) can be realized. And then, the pin-on-disk tribotest can be performed in different gas environments. The tribo-system is used to evaluate the super-low friction and wear properties of carbon based solid lubrication film in the utmost limit gas environment. The tirbo-system costs 57,000 US dollars.



SPECIFICATION:

Stainless vacuum chamber

       Rotary mechanical pump (gas pressure <100 Pa)


Gas environment control system

       Gas species: Oxygen, Nitrogen, Argon, Helium, Air, et al.

       Gas environment scope: 1 ppm ~100 %RH (humidity), 1 ppm~100 vol.% (Oxygen content)

       Gas environment control technique: gas cleaner (ppm-level); Gas hose (%)


Gas environment detect system

       Measurement range: <1 ppm~100 %RH (humidity), 1 ppm~100 vol.% (Oxygen content)

       Measurement method: Sensors in the gas cleaner (ppm-level), Oxygen content analysis meter and humidity probe


Pin-on-Disk Tribotest system

       Rotary motion and reciprocate movement, normal load 0.1~2 N